Shutting down
OPERATION MANUAL
Chapter 8
Heat Transfer Station eo
V2.3.0en/22.04.20//17.12
78
8.3
Draining the cooling water
This section must be observed when using water-cooled temperature control units.
8.3.1
Draining process
Pressurized cooling water connections
RISK OF INJURY
Wear your personnel protective equipment (e.g. safety goggles).
Carefully open the cooling water connection. Open slowly (1-2 signal edges) and drain the
cooling water slowly.
Building side isolating valves are not closed
DAMAGE BY ROOM FLOODING
Close the building’s isolating valves in the cooling water supply and return lines.
PROCEDURE
Close the building’s isolating valves in the cooling water supply and return lines.
Place a collecting container below the
>Cooling water inlet<
[13] and
>Cooling water out-
let<
[14].
Open the
>Cooling water inlet<
[13] and
>Cooling water outlet<
[14]. The cooling water will
begin to drain out. Allow all the cooling water to drain out to prevent the risk of freezing during
transport and storage!
8.4
Deinstalling the collecting container
PROCEDURE
Remove the hose from the collecting container.
Follow the instructions for the disposal of thermal fluid on page 15 in Section
.
Dismantle the hose at the
>Overflow<
[12].
8.5
Packing
Use the original packaging wherever possible! Further information can be found on page 22 in sec-
tion
.
8.6
Shipping
Temperature control unit transported in a horizontal position
DAMAGE TO THE COMPRESSOR
Only transport the temperature control unit in an upright position.
Temperature control unit transported incorrectly
PROPERTY DAMAGE
Do not transport by truck on the castors or feet.
Comply all requirements in this section to avoid damage to the temperature control unit.