Function description
Chapter 3
OPERATION MANUAL
V2.3.0en/22.04.20//17.12
Heat Transfer Station eo
41
In the following dialog, enter the overtemperature protection suitable to the thermal fluid used.
To reset the cut-out values of the overtemperature protection to the factory setting from within
the controller of Pilot ONE, enter for “Heater” 35 °C. “Process Safety” is factory set to “Stop” and is
automatically reset to “Stop” when restoring the factory settings.
Enter the temperature unit you want to use with Pilot ONE. The available choices are “Celsius
(°C)”, “Kelvin (K)” and “Fahrenheit (°F)”.
Confirm your choice by tapping on “OK”.
Read the message and confirm by tapping on “OK”.
Read the safety warning and confirm by tapping on “OK”.
Read the Note and confirm by tapping on “OK”.
Tap on the temperature format set by you in the controller (green text).
Enter the displayed red safety code using the numeric keypad that appears.
Ente
r the value 35 °C for the “Heating OT value” using the numeric keypad that appears.
Confirm your entry by tapping on “OK”. The message “Restart system!” appears on the
>Touchscreen<
[88].
Switch off the temperature control unit. The temperature control unit is reset.