Setup mode
OPERATION MANUAL
Chapter 4
Heat Transfer Station eo
V2.3.0en/22.04.20//17.12
58
ly. If tapping on “OK” is not correct, it will be displayed graphically for 2 seconds. The display will
return to the “Home” screen. Try starting the thermoregulation again.
Carry out temperature control to the entered setpoint until no more gas bubbles rise up.
Increase the setpoint by 10 K and carry out temperature control until no more gas bubbles rise
up.
Repeat increasing the setpoint by 10 K until the maximum working temperature of the thermal
fluid has been reached.
Tap on the “Stop” touch button as soon as no more gas bubbles rise up at the thermal fluid's
maximum working temperature.
Confirm the stop of the temperature control process by tapping on “OK”.
The correct selection will be displayed graphically. Thermoregulation is stopped immediately and
the pump continues to run for approx. 30 seconds. Wait until the pump stops. If tapping on “OK”
is not correct, it will be displayed graphically for 2 seconds. The display will return to the “Home”
screen. Try stopping the temperature control process again. The de-gassing process is complete.
4.3.1.3
Draining externally open applications
Hot or very cold thermal fluid
SEVERE BURNS/FROSTBITE OF LIMBS
Before draining, ensure that the thermal fluid has room temperature (20 °C).
If, at this temperature, the thermal fluid is too viscous to be drained: Control the temperature
of the thermal fluid for a few minutes until the viscosity will allow drainage. Never control the
temperature of the thermal fluid when the Drain is open.
Danger of burns when draining thermal fluids at temperatures above 20
°C.
Wear your Personal Protective Equipment for draining.
Only drain using suitable drainage hose and collecting container. These must be compatible
with the thermal fluid and its temperature.
PROCEDURE
Have a suitable container ready to catch the thermal fluid.
Empty the externally open application as described in the accompanying operating manual. Wait
until the externally open application is empty. Check if the thermal fluid can be reused. On page
15 observe section
»Proper disposal of resources and consumables«
.
Open the knurled screw at the rear
>Drain<
[10]. The thermal fluid will flow from the tempera-
ture control unit into the container as soon as you have opened the knurled screw. Wait until no
more thermal fluid flows from the temperature control unit. Check if the thermal fluid can be re-
used. On page 15 observe section
»Proper disposal of resources and consumables«
.
Open the knurled screw at the front
>Drain<
[10]. The thermal fluid will flow from the tempera-
ture control unit into the container as soon as you have opened the knurled screw. Wait until no
more thermal fluid flows from the temperature control unit. Check if the thermal fluid can be re-
used. On page 15 observe section
»Proper disposal of resources and consumables«
.
Open the connection
>Circulation flow<
[1].
Open the connection
>Circulation return<
[2].
Leave the temperature control unit open for a while for the residue to fully drain and to allow it
to dry out (without screw caps).
Close the connection
>Circulation flow<
[1].
Close the connection
>Circulation return<
[2].
Re-fit the knurled screw at the rear and front
>Drain<
[10].