Introduction
OPERATION MANUAL
Chapter 1
Heat Transfer Station eo
V2.3.0en/22.04.20//17.12
18
1.5
Exemplary illustrations of the cooling variants
1.5.1
Consequence of inadequate energy dissipation
Room air/cooling water
Consequences of, for instance, contamination of the liquefier fins, inadequate clearance between
temperature control unit to wall/bath wall, room air/cooling water too warm, cooling water differ-
ential pressure too low, suction strainer contamination: The refrigerant in the coolant circuit can no
longer fully discharge the admitted energy to the room air/cooling water. Thus there is not sufficient
liquefied refrigerant available, the condensation temperature and the energy consumption to rise.
Example: Air and
water cooling
Air cooling: Air inlet
Water cooling: Water
connection