Setup mode
OPERATION MANUAL
Chapter 4
Heat Transfer Station eo
V2.3.0en/22.04.20//17.12
56
In the display that follows, confirm your entry again by tapping on “OK”.
The correct selection will be displayed graphically and the setpoint will be changed immediately.
If tapping on “OK” is not correct, this is displayed graphically for 2 seconds. After this, the display
will return to the “Home” screen again. Try changing the setpoint again.
4.3
Filling, venting, degassing and draining
The illustration “connection diagram” can be found on page 81 in section
.
Extremely hot / cold surfaces, connections and thermal fluids
BURNS/FREEZING OF LIMBS
Surfaces, connections and tempered thermal fluids can be extremely hot or cold depending on
the operating mode.
Avoid direct contact with surfaces, connections and thermal fluids!
Wear your personnel protective equipment (e.g. temperature-resistant safety gloves, safety
goggles).
During an active circulation, the thermal fluid circuit is shut off by shut-off valves
MATERIAL DAMAGE TO THE CIRCULATING PUMPS INSTALLED IN THE TEMPERATURE CONTROL
UNIT
Do not close the thermal fluid circuit during an active circulation by means of shut-off valves.
Warm the thermal fluid to room temperature before stopping the circulation.
4.3.1
Externally open application
4.3.1.1
Filling and venting externally open application
Non-compliance with the safety data sheet for the thermal fluid to be used
INJURIES
Risk of injury to the eyes, skin, respiratory tract.
The safety data sheet for the thermal fluid to be used must be read prior to using it and its
content must be respected.
Observe the local regulations/work instructions.
Wear your personal protective equipment (e.g. temperature-resistant safety gloves, safety
goggles, safety footwear).
Danger of slipping because floor and work area are contaminated. Clean the work station and
follow the instructions for the disposal of thermal fluid and material on page 15 in section
»Proper disposal of resources and consumables«
.
▪
During the fill process, ensure any necessary measures, such as earthing the tanks, funnels and
other aids, have been taken.
▪
Fill to the lowest possible height.
PROCEDURE
Manually close the knurled screw at the
>Drain<
[8].
Check that a hose was installed at the
>Overflow<
[12] (if present). The other end of the hose
must be inserted in a suitable collecting container. Excess thermal fluid will leak at this point
when the temperature control unit is overfilled. Hose and container must be compatible with the
thermal fluid and its temperature.
Slowly
fill suitable thermal fluid into the open application. The thermal fluid flows via the
>Circu-
lation return<
[2] into the temperature control unit and via the
>Circulation flow<
[1] back to
the externally open application. Thereby the air in the temperature control unit reaches the out-
side. Follow the instructions for cleaning the filling accessories on page 15 in section
disposal of resources and consumables«
.
Switch on the temperature control unit.
Go to the “Categories Menu”.
Tap on the category “Temperature Control”.
Tap on the category “Start/Stop”.
Tap on the dialog entry “Start venting”.
Confirm your choice by tapping on “OK”.