Interfaces and software update
OPERATION MANUAL
Chapter 6
Heat Transfer Station eo
V2.3.0en/22.04.20//17.12
66
Note that you may only change
>Com.G@te<
[46] when the temperature control unit is turned off.
In the category “Interfaces” you can change the settings of each function such as PoKo, analog
interface and RS232/RS485.
You may order
>Com.G@te<
[46] with Order No. 6915. You can extend an existing
>Com.G@te<
[46] via a connecting cable (Order No. 16160).
6.3.1
Jack LEVEL (Com.G@te® external only)
For level monitoring in the
>Sight glass<
[23].
This connection enables you to connect an external float switch (Order No. 6152), which is posi-
tioned in the
>Sight glass<
[23], for monitoring the level of your externally closed application. Acti-
vation via a potential-free contact.
The interface is specified as a digital input. Do not apply voltage or current.
Pin
Description
1
Level test (bypass via contact 2 → “absence”)
2
Level – (GND)
3
Level + (normally open contact)
6.3.2
Connector POKO (floating contact) alarm
Information on the interface can be found on page 63 in Section
»Connector POKO (floating con-
.
6.3.3
Jack AIF Reg-E-Prog
The analog interface has a programmable input channel and 3 output channels.
The analog interface of Com.G@te® is programmed in the category “Interfaces”.
Pin
Description
Signal
1
Current output, T external
0
a)
/4 - 20 mA or 0 - 10 V
If using 0 - 10 V, integrate a
500 Ω
resistor.
2
Current output, setpoint
0
a)
/4 - 20 mA or 0 - 10 V
3
GND for analog outputs
GND
4
Analog input (programmable)
0
a)
/4 - 20 mA or 0 - 10 V
Power input: 200 Ω working
resistance
Voltage input: 100 kΩ input
resistance
Pin assignment (front
view)
Pin assignment
Pin assignment (front
view)
Pin assignment