4-3.
SWEEP TIME ACCURACY (Cont’d)
3.
Set up the 5316A as follows:
TIA—~B
IN
GATE TIME
MIN
LEVEL/SENSE (CHANNEL A and B)
+MAX
TRIGGER (CHANNEL A and B)
LEVEL
AC/DC (CHANNEL A and B)
DC
ATTN (CHANNEL A and B)
Xl
FILTER
NORM
SEP/COM A
COM A
Channel A Triggering
Leading Edge
Channel B Triggering
Trailing Edge
NOTE
To ensure that an accurate sweep time indication is obtained from
the HP 5316A in step 4, allow the DUT to complete at least three
sweeps.
4. On the HP 8340B/41B, press [SWEEP TIME] [1] [0] [msec]. Verify that
the sweep time
displayed by the HP 5316A is within the limits shown in Table 4-6.
5. Repeat step 4 for values shown in Table 4-6.
Table 4-6.
Sweep Time Accuracy Limits
Selected
Lower Limit
Upper Limit
10 ms
9.S ins
10.S ms
lOOms
9Sms
lOSms
is
.9Ss
l.OSs
lOs
9.Ss
10.Ss
50 s
47.S s
52.S s
4-20
Performance Tests
HP 8340B/41B