5-4.
MIN LOOP, A32, A33, AND A32A1 (Cont’d)
2. Set LINE switch to ON and press [INSTRPRESET]. Connect a jumper between
A59TP4 OLl and
5V to disable the UNLK indicator circuit.
3. Press [CW] on the OUT, then enter [6] [0] [9] [0] [MHz].
4. Disconnect the cable from A33J2 M/N OUT and connect this output to Spectrum Analyzer RF
INPUT. Set the Spectrum Analyzer as follows:
CENTER FREQ
197.419 MHz
REF LEVEL
3dBm
ATTEN
+20dB
LOG SCALE
.
. 1 dB/DIV
RES BW
.
..
300 Hz
VBW
300 Hz
FREQ SPAN
5 kHz
SWEEP TIME
1 SEC.
MARKER ENTRY
Press PEAK SEARCH
MARKER ENTRY
..
Press MKR
—~
CF
MARKER MODE
Press SIGNAL TRACK
5. The M/N output frequency indicated on Spectrum Analyzer should be 197.419 MHz o1 count.
NOTE
To display the frequencies that the processor programs, press [SHIFT]
[Ml] on the DUT. It will display from left to right: the M divide number, the
N divide number, the M/N output frequency, and the 20/30 output fre-
quency.
BOTTOM VIEW
A32/A33
M/N VGO AND
OUTPUT
EZZI
0
A59
7
A32133
A33J1
355-395 MHz
OUT
A33J2
A32A2C1
M/N OUT FREC ADJ
Figure 5-8.
Location of M/N Loop Adjustments
A32TP1
TUNE
A32A2C5
PWR
A33J3
5-24
Adjustments
HP 8340B/41B