HP 8340B/41B Test Record Card (24 of 30)
SPECIFICATIONS TESTED
Limits
Step
TEST
Conditions
Lower
Limit
Measured
Value
Upper
Limit
4-13. Pulse Modulation Accuracy
(Cont’d)
NOTE
Pulse Repetition Freq. Range
8340B CW Freq. = 22 GHz
10
Perform the following steps only
when testing an HP 8340B.
Minimum PRF
PRF = 100 Hz
PW = 1 microsecond
6
ENTRY DISPLAY Power Level
dBm
7
ENTRY DISPLAY Power Level
dBm
8
A Power between steps 6 and 7
>—0.3 dB
dB
<+0.3dB
Maximum PRF
PRF = 5 MHz
PW = 100 nanoseconds
6
ENTRY DISPLAY Power Level
dBm
7
ENTRY DISPLAY Power Level
dBm
8
A Power between steps 6 and 7
>—0.3 dB
dB
<--1.5 dB
4-112
Performance Tests
HP 8340B/41B