Section V.
Adjustments
INTRODUCTION
This section provides adjustment procedures for the HP 8340B/41B Synthesized Sweepers. These
procedures should not be performed as routine maintenance but should be used (1) after replacement
of a part or component, or (2) when performance tests show that the instrument’s specifications
cannot be met. Before attempting any adjustment, allow 1 hour warm-up time for
the
instrument.
Table
S-i lists the
adjustment procedures by test number in the order they appear in this Section. Table 5-2
lists all factory selected
components.
Table
5-3 lists adjustment procedures that interact between
assemblies. This table lists the adjustment paragraphs that must be checked when an assembly is
adjusted, parts replaced, or the assembly replaced.
SAFETY CONSIDERATIONS
Although this instrument has been designed in accordance with international safety standards, this
manual contains information, cautions, and warnings
which must be followed to ensure safe opera-
tion and to retain the instrument in safe condition. Service and adjustments should be performed only
by qualified service personnel.
~NG
Adjustments in this section are performed
with power supplied to the instrument while
protective covers are removed. There are voltages at many
points in the instrument
which can, if contacted, cause personal injury or death. Be extremely careful. Adust-
ments should be performed only by trained service personnel.
Power is still applied to this instrument with the LINE
switch set to STANDBY. There is no
OFF position on the
LINE switch. Before removing or installing any assembly or printed
circuit board, remove the power cord from the rear of the instrument.
Capacitors inside
the instrument may still be charged, even
if the instrument has been
disconnected from
its source of ac power. Use a non-metallic adjustment tool whenever
possible.
EQUIPMENT REQUIRED
Table 4-1
lists the equipment required for the adjustment procedures. If the test equipment recom-
mended is not available, other equipment may be used if its performance meets the “Critical Specifi-
cations” listed in the table. The test setup used for an
adjustment
procedure is referenced in each
procedure.
HP 8340B/41B
Adjustments
5-1