4-16.
FM ACCURACY AND FLATNESS (Cont’d)
HP 83408/415
UNDER TEST
SPECTRUM ANALYZER
Procedure
1. Connect the equipment as shown in Figure 4-20, FM Accuracy Setup. The BNC Tee should be
connected directly to the DUT’s FM input BNC.
2. Set the function generator to a 100 kHz sine wave. Set the DVM to ac volts.
3. Press [INSTRPRESET] on the DUT and set up the following conditions:
Sweep Mode:
CW
Frequency out:
2.2 GHz
Leveling:
NT
RF Output:
0 dBin
RF Peaking (PEAK):
ON
FM
ON
FM
Sensitivity:
1 MHz/V (default condition)
Figure 4-20.
FM Accuracy Setup
4-76
Performance Tests
HP 8340B/41B