5-10.
YO DELAY COMPENSATION, A54 (Cont’d)
Procedure
1. Position the OUT in the Test Position as shown in Figure 5-28 with bottom cover removed. Turn
the LINE switch on. Connect equipment as shown in Figure
5-28 and allow one-half hour warm-up
time.
YO KICK PULSE ADJUSTMENT
2. On the OUT, make the following settings:
a.
Press [INSTRPRESET]
b.
Press [STARTFREQ] then enter [2]
[.]
[3] [0Hz]
c.
Press [STOPFREQ] then enter [6]
[.]
[9] [9] [9] [0Hz].
3. Adjust A54R36 (PW, Figure 5-30) for a 12.5 insec pulse on the oscilloscope.
Z—AX I
DUT
ii
Figure 5-29.
YO Delay Compensation Test Setup
5-58
Adjustments
HP 8340B/41B