Hardware
Broken
Errors
7000
{
7999
7044
YTO
tuning
range
YIG-tuned
oscillator
tuning
range
Cause:
The
YIG
oscillator
is
not
able
to
be
tuned
over
the
necessary
range
.
This
error
will
only
occur
on
power
up
of
the
system,
or
when
the
TEST
command
is
invoked,
either
from
the
front
panel
or
remotely
.
Cure:
Return
to
Hewlett-P
ackard
for
servicing.
7045
10
MHz
out
error
The
HP
70310A
precision
frequency
reference
10
MHz
OUT
signal
is
absent
or
too
low
.
The
most
probable
cause
is
faulty
HP
70310A
precision
frequency
reference
hardware
.
7046
Illegal
bandslct
cde
Illegal
band-select
code
Cause:
The
Internal/External
Frequency
reference
switch
was
set
to
an
illegal
position
Cure:
Please
document
all
steps
taken
that
led
to
this
error
condition
and
contact
your
HP
representative
.
7047
RAM
failure
Cause:
The
RAM
self
test
detected
a
failure
in
the
system
RAM.
Cure:
Return
to
Hewlett-P
ackard
for
servicing.
7048
FFS
won't
unlock
Cause:
The
module
self
test
tunes
the
synthesizer
to
a
frequency
outside
it's
operating
range
to
verify
the
lock
detector
is
operating
properly
.
The
synthesizer
never
indicated
an
unlock
condition.
Cure:
Return
to
Hewlett-P
ackard
for
servicing.
7050
Hardware
config
fail
Cause:
On
power
up
of
the
system
a
test
of
the
hardware
is
made
to
verify
the
hardware
matches
the
model
revision
(\A"
or
\B").
If
this
test
is
not
conclusive
,
this
error
will
occur
.
Cure:
If
an
external
frequency
reference
is
being
used,
verify
the
proper
accuracy
.
Return
to
Hewlett-P
ackard
for
servicing.
6-32
Error
Messages