![HP 70000 series User Manual Download Page 317](http://html.mh-extra.com/html/hp/70000-series/70000-series_user-manual_161488317.webp)
User
Application
Errors
0001
{
0999
0053
Res
BW
--
CF
ampl
Cause:
The
resolution
bandwidth
of
one
or
more
of
the
HP
70903A
IF
sections
is
out
of
adjustment.
Cure:
P
erform
the
HP
70903A
IF
section
Resolution
Bandwidth
Filter
adjustments
.
0060
RF
attenuator
Cause:
The
RF
attenuator
of
the
HP
70600A
preselector
or
the
HP
70601A
preselector
is
not
functioning
correctly
.
Cure:
Check
and
replace
the
RF
Attenuator
in
the
HP
70600A
preselector
or
HP
70601A
preselector
.
0061
Cannot
test
70905/6B
Cause:
These
RF
sections
can
only
be
tested
with
an
associated
HP
70600A
preselector
or
an
HP
70601A
preselector
,
as
part
of
a
preselected
microwave
spectrum
analyzer
system.
0062
Cal
input
incorrect?
Cause:
The
front
panel
CALIBRA
TOR
output
of
the
HP
70900B
local
oscillator
source
is
not
connected
to
the
RF
INPUT
of
the
system
being
tested.
(This
is
a
prerequisite
for
running
the
program.)
Cure:
This
error
indicates
that
the
marker
reading
of
the
signal
was
less
than
060
dBm
and
the
21.4
MHz
diagnostic
detector
indicated
a
low
signal
condition.
Check
that
the
calibrator
signal
from
the
HP
70900B
local
oscillator
source
is
connected
to
the
RF
Input
of
the
RF
section
or
the
preselector
.
If
the
calibrator
is
connected,
check
the
RF
signal
path
in
the
RF
section
for
excessive
loss
.
0063
ATTENUATOR
Section
Cause:
One
or
more
of
the
RF
attenuator
sections
in
the
HP
70600A
preselector
or
the
HP
70601A
preselector
is
not
functioning
correctly
.
Cure:
Check
and
replace
the
RF
Attenuator
in
the
HP
70600A
preselector
or
HP
70601A
preselector
.
6-10
Error
Messages