Hardware
Broken
Errors
7000
{
7999
Hardware
Broken
Errors
7000
{
7999
Hardware-broken
errors
occur
when
the
spectrum
analyzer
might
have
faulty
hardware
.
7000
ROM
Check
error
Cause:
The
programmed
checksum
of
the
ROM
does
not
agree
with
the
computed
checksum.
Cure:
This
could
only
be
caused
by
a
hardware
failure
.
Return
to
Hewlett-P
ackard
for
servicing.
7001
LO
unleveled
7002
First
LO
unleveled
7003
Second
LO
unlocked
7004
300
MHz
error
7005
321.4
MHz
error
7006
21.4
MHz
error
7007
Cal
error
Calibration
error
.
7008
FFS
handshake
Fractional-frequency-source
handshake
error
Cause:
Communication
has
been
lost
to
the
Frequency
Synthesizer
portion
of
the
70900
module
.
This
is
very
likely
due
to
a
hardware
failure
.
Cure:
Return
to
Hewlett-P
ackard
for
servicing.
Error
Messages
6-25