Hospira GemStar System Operating Manual
85
Chapter 6. Continuous
GemStar
Continuous Programming Worksheet
GIVE:
*Air Sensitivity may not show in all setting configurations.
Parameter
Selections and/or Values to be
Entered
Programmable Ranges
Therapy
Continuous
N/A
Unit of Measure
Concentration:
mg/mL
Concentration:
mcg/mL
mL
0.1 - 1000 mg/mL
1 - 1000 mcg/mL
Rate
0.1 - 9999 mg/hr
1 - 999900 mcg/hr
0.1 - 1000 mL/hr
VTBI
(Volume To Be
Infused)
0.1 - 99999 mg
1 - 9999000 mcg
0.1 - 9999 mL
KVO
(Keep Vein
Open)
No
KVO Rate: 0.0
Yes
KVO Rate:
0.1 - 5.0 mL/hr
Container Size
(VTBI + KVO)
0.1 - 99999 mg
1 - 9999000 mcg
0.1 - 9999 mL
Air Sensitivity*
On
Alarms at approx. 0.5 mL of air
2 mL Alarms at approx. 2 mL of air
Off
No alarm
N/A
Piggyback Dose
Amount
1 - 7200 mL
Piggyback
Infusion Time
hrs:min
1 min or
dose amount / 300 mL/hr to 24
hrs or dose amount / 0.1 mL/hr
Summary of Contents for GemStar Series
Page 9: ...Notes viii Hospira GemStar System Operating Manual ...
Page 27: ...Notes 18 Hospira GemStar System Operating Manual ...
Page 67: ...Notes 58 Hospira GemStar System Operating Manual ...
Page 77: ...Notes 68 Hospira GemStar System Operating Manual ...
Page 87: ...Notes 78 Hospira GemStar System Operating Manual ...
Page 95: ...Notes 86 Hospira GemStar System Operating Manual ...
Page 103: ...Notes 94 Hospira GemStar System Operating Manual ...
Page 165: ...Notes 156 Hospira GemStar System Operating Manual ...
Page 173: ...Chapter 14 Operation Test 164 Hospira GemStar System Operating Manual ...
Page 175: ...Notes 166 Hospira GemStar System Operating Manual ...
Page 193: ...Notes 184 Hospira GemStar System Operating Manual ...
Page 221: ...Notes 212 Hospira GemStar System Operating Manual ...