Chapter 9. Variable Time
106
Hospira GemStar System Operating Manual
Changing a Variable Time Program
To change a program:
1
From the
STOP
mode, press [
CHANGE
] and select
CHANGE PROGRAM
.
2
As each program parameter displays, make the desired changes or press
[
YES/ENTER
] to accept the current setting.
3
Review the program. Press
to begin review.
4
Press [
START
] to begin delivery.
Can be changed at
any time
Can ONLY be changed
before
is pressed
Cannot be changed
•
KVO rate
•
Container size
•
Air sensitivity
•
Bolus
•
Current time
•
Base rate
•
Daily amount
•
Phase parameters
•
Unit of measure
•
Concentration
•
Phase type
d
Note: You cannot change a program when the pump is in
FULL
or
CONTAINER
Lock.
d
Note: The clock cannot be changed while a Variable Time Program is in progress.
If
[START]
has already been pressed in a Pain Management program, when the SET
CLOCK function is selected from Stop Mode, the following message shall display from
2 - 5 seconds during which no keys shall be valid except
[YES/ENTER]
as a quick
escape, before returning to the Stop Mode display:
d
Note: The container size cannot be made smaller than the daily total of phases, base
rate, and KVO combined; or the current amount delivered, whichever is greater.
Summary of Contents for GemStar Series
Page 9: ...Notes viii Hospira GemStar System Operating Manual ...
Page 27: ...Notes 18 Hospira GemStar System Operating Manual ...
Page 67: ...Notes 58 Hospira GemStar System Operating Manual ...
Page 77: ...Notes 68 Hospira GemStar System Operating Manual ...
Page 87: ...Notes 78 Hospira GemStar System Operating Manual ...
Page 95: ...Notes 86 Hospira GemStar System Operating Manual ...
Page 103: ...Notes 94 Hospira GemStar System Operating Manual ...
Page 165: ...Notes 156 Hospira GemStar System Operating Manual ...
Page 173: ...Chapter 14 Operation Test 164 Hospira GemStar System Operating Manual ...
Page 175: ...Notes 166 Hospira GemStar System Operating Manual ...
Page 193: ...Notes 184 Hospira GemStar System Operating Manual ...
Page 221: ...Notes 212 Hospira GemStar System Operating Manual ...