Chapter 5. Intermittent
72
Hospira GemStar System Operating Manual
d
Note: The call back alarm works only when the pump is powered on AND in the
STOP
mode. DO NOT power-off the pump while it is disconnected from the patient
if you have set the call back alarm.
When the alarm sounds, press
[
SILENCE
]
to silence the alarm for three minutes.
Then reconnect the pump to the patient.
f
Tip: To use a delayed start without specifying a start time:
1
Press [
YES/ENTER
].
2
Complete the program review.
3
Power-off the pump.
4
When the pump is powered on again, complete the program review, then
press [
START
] to begin delivery.