Chapter 2. Basic Operation
52
Hospira GemStar System Operating Manual
12
Press [
PURGE
].
13
The pump displays
PRIME THE
SET?
Press [
YES/ENTER
].
14
Press and hold [
PURGE
].
Continue using the [
PURGE
] key
until fluid fills the tubing. Ensure
all air is removed from the
container, cassette, and tubing.
CAUTION: When priming is
complete, ensure no fluid flows
at the distal end of the pump set.
If flow is observed,
DO NOT use the pump set.
15
Connect the pump set to the
patient-access device.
d
Note: Purge rate is 250 mL/hr.
The pump purges for up to 2
minutes each time you press and
hold
[
PURGE
]
.