Hospira GemStar System Operating Manual
133
Chapter 11. Clinician Instructions
Adding a Second Lock Level
1
Press [
OPTIONS
] to display the Options Menu.
2
Select
KEYPAD LOCK
from the Options Menu.
3
Enter the Lock Sequence Number.
4
For
RATE CHANGE
Lock, set the minimum and maximum programmable ranges
for the continuous delivery rate and bolus dose in a Pain Management program.
•
MIN
flashes on the display. Enter the minimum value.
•
MAX
flashes on the display. Enter the maximum value.
5
Press [
YES/ENTER
] when done.
6
Repeat steps 4 and 5 for a bolus dose if available in the current program.
7
Enter the lock sequence number. Asterisks display as the numbers are entered.
•
For
THERAPY
Lock, enter [
0
][
0
][
0
][
3
][
1
].
•
For
UNLOCKED
,
FULL
,
CONTAINER
, and
RATE CHANGE
lock levels, enter
[
1
][
3
][
0
][
0
][
0
].
f
Tip: Press
to toggle between the MIN and MAX values.
d
Note: When two lock levels are applied, if a function is available under one lock
level but not the other, the more stringent restrictions apply.
d
Note: Default lock sequence numbers are shown.
Summary of Contents for GemStar Series
Page 9: ...Notes viii Hospira GemStar System Operating Manual ...
Page 27: ...Notes 18 Hospira GemStar System Operating Manual ...
Page 67: ...Notes 58 Hospira GemStar System Operating Manual ...
Page 77: ...Notes 68 Hospira GemStar System Operating Manual ...
Page 87: ...Notes 78 Hospira GemStar System Operating Manual ...
Page 95: ...Notes 86 Hospira GemStar System Operating Manual ...
Page 103: ...Notes 94 Hospira GemStar System Operating Manual ...
Page 165: ...Notes 156 Hospira GemStar System Operating Manual ...
Page 173: ...Chapter 14 Operation Test 164 Hospira GemStar System Operating Manual ...
Page 175: ...Notes 166 Hospira GemStar System Operating Manual ...
Page 193: ...Notes 184 Hospira GemStar System Operating Manual ...
Page 221: ...Notes 212 Hospira GemStar System Operating Manual ...