Hospira GemStar System Operating Manual
73
Chapter 5. Intermittent
Intermittent Delivery Interruptions
Missed Start Time
6
If the programmed start time is missed, it can be changed through the Options
Menu. Follow the steps below to reprogram the start time.
6
If the pump remains in the
STOP
mode and the first dose is missed, press
[
START
] to begin delivery immediately and all remaining doses will be shifted.
6
If the pump is programmed with a delayed start time and is then powered off
before [
START
] is pressed, the pump will skip to the NEXT occurrence of the
programmed start time.
For example, the current time is 10:00 PM. The user sets the delayed start time for
10:30 PM and then powers off the pump. The user does not power on the pump
until 10:40 PM, which is 10 minutes AFTER the scheduled start time. The pump
does not begin delivery until 10:30 PM the FOLLOWING day.
Reprogramming Start Time
1
Press [
OPTIONS
] and select
NEXT DOSE
.
2
Use the number keys to enter the start time for the
next dose. Press [
YES/ENTER
] to accept the change
and exit
OPTIONS.
d
Note: The clock cannot be changed while an Intermittent Program is in
progress.
WARNING: If
NO START DELAY
is seen onscreen after the program is
confirmed, the first dose will begin infusing immediately after
the
[START]
button is pressed. If this is not desired, the
infusion start time may be reprogrammed. Follow the steps
below to reprogram the start time.
f
Tip: When entering the time in 12-hour format, press
or
to toggle between
AM and PM.
1
REVIEW PROGRAM
2
HISTORIES
3
KEYPAD LOCK
4
PUMP SETTINGS
5
SET CLOCK
6
7
SPEED PROTOCOL
8
NEXT DOSE
Summary of Contents for GemStar Series
Page 9: ...Notes viii Hospira GemStar System Operating Manual ...
Page 27: ...Notes 18 Hospira GemStar System Operating Manual ...
Page 67: ...Notes 58 Hospira GemStar System Operating Manual ...
Page 77: ...Notes 68 Hospira GemStar System Operating Manual ...
Page 87: ...Notes 78 Hospira GemStar System Operating Manual ...
Page 95: ...Notes 86 Hospira GemStar System Operating Manual ...
Page 103: ...Notes 94 Hospira GemStar System Operating Manual ...
Page 165: ...Notes 156 Hospira GemStar System Operating Manual ...
Page 173: ...Chapter 14 Operation Test 164 Hospira GemStar System Operating Manual ...
Page 175: ...Notes 166 Hospira GemStar System Operating Manual ...
Page 193: ...Notes 184 Hospira GemStar System Operating Manual ...
Page 221: ...Notes 212 Hospira GemStar System Operating Manual ...