2. Command
specifications
2.1. Equipment operating console
The equipment is operative only by the SELECT key.
Receiving the communication commands described in 2.2 below from the host computer (CPU),
the equipment controls itself according to the commands received.
It also returns the response data to the host computer.
2.2. Communication commands (CPU to SF)
Command
Format Delimiter
Command
function
description
E
"E"
<CR>
Similar when pressing SELECT key and START key
simultaneously, equipment aborts normal measurement,
automatic measurement and automatic wetting
operation.
R
"R"
<CR>
Similar when pressing START key for one second or
longer, equipment carries out automatic wetting
operation.
S
"S1"
<CR>
Similar when pressing START key, equipment carries out
normal measurement.
"S0"Other thanS2
〜
10
<CR>
"S2"
<CR>
Automatic measurement twice Similar when pressing
START key in AUTO
mode, equipment
carries out automatic
measurement.
"S3"
<CR>
Automatic measurement 3times
⋮
⋮
⋮
⋮
⋮
⋮
"S10" <CR>
Automatic
measurement
10times
P
"P730.0"
<CR>
Sets to minimum atmospheric pressure (730.0 hPa).
"P1000.0"
P730.0
〜
P1070.0
<CR>
Sets to atmospheric pressure between minimum and
maximum.
"P1070.0"
<CR>
Sets to maximum atmospheric pressure (1070.0 hPa).
"P1070.1"
P1070.1 or greater
<CR>
If a value greater than maximum atmospheric pressure
or lower than minimum is commanded, equipment
ignores it and sets to 1013.3 hPa.
"P729.9"
Less than P730.0
<CR>
"P0.0"
<CR>
Equipment sets measured atmospheric pressure to itself.
(Only for SF-2)
Summary of Contents for SF-1U
Page 1: ...FILM FLOW METER SF 1U 2U Instruction Manual CODE I030248700E...
Page 2: ...May 2016 2004 2016 HORIBA STEC Co Ltd...
Page 3: ...J i...
Page 4: ...J ii...
Page 5: ...J iii 1 2 3 4 5 6 7 8 1 40 2 3 4 5 40 6 7 8 9 10 11 12 13...
Page 7: ...J 1 SF 1U SF 2U Snoop Swagelok Company...
Page 8: ...J 2 1 SF VP 1384428 SF 2U 2 0 35 1013 3hPa 1 3 SF 1U SF 2U 1 Snoop 59ml 1 I030248700 1 AC...
Page 10: ...J 4 5 2...
Page 12: ...J 6 6 6 1 6 1 1 6 1 2 AC...
Page 13: ...J 7 6 1 3 VP 1U 4U SF 3 4 6 1 4 2 3cm 4cm Snoop Snoop...
Page 14: ...J 8 6 1 5 VP 1U 3U 6mm 5 6mm VP 4U 10mm 9 10mm 10mm 9 10mm 3L min 2cm VP 1U 2U 1 2mm...
Page 15: ...J 9 6 1 6 25 5 0 35 25 3 6 1 7 3 10 6 2 6 2 1 N2 5 20 1 1...
Page 16: ...J 10 6 2 2 SF 1U 730 1070hPa 1013 3hPa SF 2U 2 730 1070hPa 1013 3hPa 2 6 2 3 2...
Page 19: ...J 13 8 8 1...
Page 20: ...J 14 8 2 A1 A1 A1 120 VP 1U 180 120 180 6 4 120 120 6 4 A1 7 1 3...
Page 21: ...J 15 8 3 A2 A2 A2 120 VP 1U 180 120 180 6 4 7 1 3...
Page 22: ...J 16 8 4 A3 A3 A3 7 1 3 8 5 A4 A4 A4...
Page 23: ...J 17 8 6 8 7 7 1 3...
Page 39: ...E 4 5 Description of each part Fig 2 Description of each part...