E-3
4. Specifications
Name FILM
FLOW
METER
Model SF-1U/SF-2U
Measuring unit model
VP-1U
VP-2U
VP-3U
VP-4U
Nominal volume
0.4 ml
2 ml
20 ml
200 ml
Flow rate measurement
range
0.2
〜
10 ml/min
2
〜
100 ml/min
20
〜
1000 ml/min
0.2
〜
10 L/min
Measurement time
120
〜
2.4 sec
60
〜
1.2 sec
60
〜
1.2 sec
60
〜
1.2 sec
Accuracy
±0.5
%
0.2
〜
2 ml/min
2
〜
20 ml/min
20
〜
200 ml/min
0.2
〜
1 L/min
±1.0
%
2
〜
10 ml/min
20
〜
100 ml/min
200
〜
1000 ml/min
1
〜
10 L/min
Repeatability
±0.2
%
0.2
〜
2 ml/min
2
〜
20 ml/min
20
〜
200 ml/min
0.2
〜
1 L/min
±0.5
%
2
〜
10 ml/min
20
〜
100 ml/min
200
〜
1000 ml/min
1
〜
10 L/min
Applicable gases
N
2
, Air, O
2
, H
2
, He, Ar, CO, CH
4
, C
3
H
8
and other similar gases.
Note: Not applicable to water-soluble or corrosive gases
Setting range of
atmospheric pressure
730
〜
1070hPa
(For SF-2U, at the
manual atmospheric pressure data entry mode.
)
Range of atmospheric
pressure measurement
(SF-2U)
930
〜
1040hPa Accuracy :±2.7hPa
(Measured data can be used for correction.
When used for correction, overall accuracy becomes ±0.7% from ±0.5%
and ±1.2% from ±1.0%, respectively.)
Display
LCD 6-digit digital display.
(Readings are switched among measured flow rate, measurement time,
temperature, atmospheric pressure and other data.)
Automatic measurement
function
Takes 2 to 10 measurements and calculates mean values automatically in
AUTO mode. (Previous data can be stored and called up. Data is gone
when equipment is turned off.)
Power source
DC 7V 0.5A Dedicated AC adapter
Ambient operating
temperature
5
〜
40°C
Accuracy is guaranteed between 15 to 35°C.
Ambient operating humidity Less than 80% (no condensation)
Indoor/Outdoor usage
Indoor use only
Corrective function
Data correction by temperature, atmospheric pressure and vapor pressure
to generate flow rates at calibration temperature and standard atmospheric
pressure (1,013.3 hPa).
Calibration temperature can be set to O to 35°C.
External data input/output RS-232C-compliant
Weight
Approx. 2.1kg
Approx. 2.7kg
External dimension
W140mm x H287.6mm x D200mm
W140mm x H457.6mm
x D200mm
Applicable standards
EMC DIRECTIVE
EN61326-1 class B
Summary of Contents for SF-1U
Page 1: ...FILM FLOW METER SF 1U 2U Instruction Manual CODE I030248700E...
Page 2: ...May 2016 2004 2016 HORIBA STEC Co Ltd...
Page 3: ...J i...
Page 4: ...J ii...
Page 5: ...J iii 1 2 3 4 5 6 7 8 1 40 2 3 4 5 40 6 7 8 9 10 11 12 13...
Page 7: ...J 1 SF 1U SF 2U Snoop Swagelok Company...
Page 8: ...J 2 1 SF VP 1384428 SF 2U 2 0 35 1013 3hPa 1 3 SF 1U SF 2U 1 Snoop 59ml 1 I030248700 1 AC...
Page 10: ...J 4 5 2...
Page 12: ...J 6 6 6 1 6 1 1 6 1 2 AC...
Page 13: ...J 7 6 1 3 VP 1U 4U SF 3 4 6 1 4 2 3cm 4cm Snoop Snoop...
Page 14: ...J 8 6 1 5 VP 1U 3U 6mm 5 6mm VP 4U 10mm 9 10mm 10mm 9 10mm 3L min 2cm VP 1U 2U 1 2mm...
Page 15: ...J 9 6 1 6 25 5 0 35 25 3 6 1 7 3 10 6 2 6 2 1 N2 5 20 1 1...
Page 16: ...J 10 6 2 2 SF 1U 730 1070hPa 1013 3hPa SF 2U 2 730 1070hPa 1013 3hPa 2 6 2 3 2...
Page 19: ...J 13 8 8 1...
Page 20: ...J 14 8 2 A1 A1 A1 120 VP 1U 180 120 180 6 4 120 120 6 4 A1 7 1 3...
Page 21: ...J 15 8 3 A2 A2 A2 120 VP 1U 180 120 180 6 4 7 1 3...
Page 22: ...J 16 8 4 A3 A3 A3 7 1 3 8 5 A4 A4 A4...
Page 23: ...J 17 8 6 8 7 7 1 3...
Page 39: ...E 4 5 Description of each part Fig 2 Description of each part...