E-15
8.2. A1
alarm
(Detecting start detector errors)
A1 alarm generated
Check
filming
condition
Abnormal
Abnormal
Normal
Check filming operation
Abnormal
Check measured flow rate
Normal
※
Repair
Replace VP
Normal
Check filming
Abnormal
Abnormal
liquid level
Check
alarm
time
Adjust
Normal
liquid level
Set time
Normal
Check filming liquid
Abnormal
for contamination
Replace
Normal
filming
liquid
Abnormal
Check
measuring
tube
for
contamination
Clean measuring tube
Normal
※
Check, repair or overhaul
detector
Items marked with “
※
” are services from HORIBA STEC,CO.,LTD. or your dealer.
Placing order for the services is recommended.
A1 alarm (
A1:−−−−
)
This alarm is generated if the time from film creation to its detection by the start
detector takes more than 120 seconds (180 seconds for VP-1U) due to film breakage
or attempting to measure a flow rate outside the specification range.
Wet inside measuring tube and check the flow rate attempting to measure. The
alarm time (120 or 180 seconds) can be changed.
⇒6.4
Automatic wetting operation alarm (
A1:−−−A
)
This alarm is generated if the interior walls of the measuring tube do not get wet in
120 seconds after starting the automatic wetting operation. Check the flow rate
attempting to measure and the filming liquid level. The alarm time (120 seconds) can
be changed.
⇒6.4
(Common to the A1 alarm)
Cleaning measuring tube.
⇒7.1.3
Press the START key to reset the alarm.
Summary of Contents for SF-1U
Page 1: ...FILM FLOW METER SF 1U 2U Instruction Manual CODE I030248700E...
Page 2: ...May 2016 2004 2016 HORIBA STEC Co Ltd...
Page 3: ...J i...
Page 4: ...J ii...
Page 5: ...J iii 1 2 3 4 5 6 7 8 1 40 2 3 4 5 40 6 7 8 9 10 11 12 13...
Page 7: ...J 1 SF 1U SF 2U Snoop Swagelok Company...
Page 8: ...J 2 1 SF VP 1384428 SF 2U 2 0 35 1013 3hPa 1 3 SF 1U SF 2U 1 Snoop 59ml 1 I030248700 1 AC...
Page 10: ...J 4 5 2...
Page 12: ...J 6 6 6 1 6 1 1 6 1 2 AC...
Page 13: ...J 7 6 1 3 VP 1U 4U SF 3 4 6 1 4 2 3cm 4cm Snoop Snoop...
Page 14: ...J 8 6 1 5 VP 1U 3U 6mm 5 6mm VP 4U 10mm 9 10mm 10mm 9 10mm 3L min 2cm VP 1U 2U 1 2mm...
Page 15: ...J 9 6 1 6 25 5 0 35 25 3 6 1 7 3 10 6 2 6 2 1 N2 5 20 1 1...
Page 16: ...J 10 6 2 2 SF 1U 730 1070hPa 1013 3hPa SF 2U 2 730 1070hPa 1013 3hPa 2 6 2 3 2...
Page 19: ...J 13 8 8 1...
Page 20: ...J 14 8 2 A1 A1 A1 120 VP 1U 180 120 180 6 4 120 120 6 4 A1 7 1 3...
Page 21: ...J 15 8 3 A2 A2 A2 120 VP 1U 180 120 180 6 4 7 1 3...
Page 22: ...J 16 8 4 A3 A3 A3 7 1 3 8 5 A4 A4 A4...
Page 23: ...J 17 8 6 8 7 7 1 3...
Page 39: ...E 4 5 Description of each part Fig 2 Description of each part...