Measurement
35
Transmission output
1.A burnout hold
setting
Specify the output during burnout of the transmission output 1.A
when an error occurs during measurement.
z
out.20: Outputs 21 mA at the time of burnout
z
out.4: Outputs 3.8 mA at the time of burnout
z
non: No burnout setting
Note:
A burnout contact is output only in the meas mode.
out.20,
out.4,
non
Transmission output 1.b setting
Tip
The current 1 provides two ranges named A and b. One of the A or b will be selected by an
external input. If you want to change ranges, you have to specify the E.in1 or E.in2 to be a range
select function. If the external range select is not input the Cur1.A will be a selected range for
current 1 output.
Transmission output
1.b value (zero range)
setting
Specify the value at the zero point (4 mA) of transmission output
1.b.
0.00 to
5.00
(mg/L)
Transmission output
1.b value (range span)
setting
Specify the value at the span point (20 mA) of transmission output
1.b.
0.00 to
5.00
(mg/L)
Transmission output
hold (current hold)
setting for transmission
output 1.b
Specify the transmission output value when transmission output 1.b
is in the hold mode.
z
HoLd: Holds and outputs the latest value
z
PrES: Outputs the preset value
HoLd,
PrES
Transmission output
1.b hold output value
(preset) setting
Input the preset value for current holding.
Note:
This is displayed when “Transmission output hold (current hold)
setting for transmission output 1.b” is set to “PrES”.
0.00 to
5.00
(mg/L)
Transmission output
1.b burnout hold setting
Specify the output during burnout of transmission output 1.b.
z
out.20: Outputs 21 mA at the time of burnout
z
out.4: Outputs 3.8 mA at the time of burnout
z
non: No burnout setting
Note:
A burnout contact is output only in the meas mode.
out.20,
out.4,
non
Transmission output 2 setting
Transmission output 2
value (zero range)
setting
Specify the temperature at the zero point (4 mA) of transmission
output 2.
−
10.0 to
110.0
(
°
C)
Transmission output 2
value (range span)
setting
Specify the temperature at the span point (20 mA) of transmission
output 2.
−
10.0 to
110.0
(°
C)
Transmission output
hold (current hold)
setting for transmission
output 2
Specify the transmission output value when transmission output 2 is
in the hold mode.
z
HoLd: Holds and outputs the latest value
z
PrES: Outputs the preset value
HoLd,
PrES
Display
Item
Description
Setting
range