Measurement
39
Q
Output conditions for each output type
The output conditions for various output types are shown below.
FAIL output setting
Contact
output
FAIL output
*1
Transmission output
(Cur1.A, Cur1.b)
Transmission output (Cur2)
C.HoLd
C.HoLd
non
CtrL
, AL,
t
HoL
d
CAL
HoLd
PrES
HoLd
PrES
b.out
b.out
non
*4
non
*4
non
*4
non
*4
Power OFF
OFF
Opened (ON)
-
Mea
su
rement
When normal operation
Valid
Closed (OFF)
Measurement value
ccurrence
of over-FS
F.oF
yES
ON
Opened (ON)
Latest
value
*5
Preset
value
*5
Latest value
Preset value
no
Closed (OFF)
Measurement value
E-11
F.CALE
yES
ON
Opened (ON)
Latest
value
*5
Preset
value
*5
no
Closed (OFF)
E-13
*3
yES
ON
Opened (ON)
*5
*5
no
Closed (OFF)
E-21 to 25
F.tE
yES
ON
OF
F
Opened (ON)
*5
*5
*5
*5
no
Closed (OFF)
Latest value
E-61
F.FLo
yES
Opened (ON)
*5
*5
Latest value
Preset value
no
Closed (OFF)
E-62, E-
75, E-86
F.SC
yES OFF
Opened (ON)
*5
*5
no
Closed (OFF)
E-80 to 82
yES
Opened (ON)
*5
*5
*5
*5
no
Closed (OFF)
Latest value
E-87
F.rt
yES
Opened (ON)
*5
*5
no
Closed (OFF)
E-90 to 92
Opened (ON)
*5
*5
*5
*5
Au
to
zero
calib
ra
tion
When normal operation,
E-11 to 13
ON
Closed (OFF)
Latest
value
Preset
value
Latest value
Preset value
E-21 to E-87
Calibration will be cancelled and the measurement mode will start if the Errors
occur.
E-90 to 92
ON
Opened (ON)
Latest
value
Preset
value
Latest value
Preset value
Ho
ldi
ng
af
te
r
cali
bration
When normal operation,
E-11 to 13
OFF
Closed (OFF)
E-21 to E-87
Calibration will be cancelled and the measurement mode will start if the errors
occur.
E-90 to 92
ON
Opened (ON)
Latest
value
Preset
value
Latest value
Preset value
External
inpu
t
HOLD
When normal operation,
E-11 to 13
Va
lid
Closed (OFF)
E-90 to 92
Opened (ON)
Hol
d
mode
*2
When normal operation,
E-11 to 13
OF
F Closed (OFF)
E-90 to 92
Opened (ON)