Measurement
27
Measurement
Meas mode
1.
Turn the power ON.
The measurement target appears on the main display.
The measurement range is then displayed and the system enters the meas mode.
After the measured value is displayed, measurement starts.
This is the normal meas mode.
2.
The screen is changed by pressing
S
/
T
.
Pressing ENT allows you to check or change the set values during measurement.
To return to the meas mode from a different mode, hold down HOLD or ESC.
Note
z
Prior to measurement, be sure to calibrate the detector. Refer to “ Calibration ” (page 53).
z
The measurement range of the residual chlorine is 0.00 mg/L to 3.00 mg/L.
The displayed value will be fixed to “3.00” and blinking when the range cut is set to “yES”.
When the range cut is set to “no”, if the measured value exceeds 5.00 mg/L, a blinking measured
value is displayed.
z
The measurement range of the temperature is 0.0
°
C to 50.0
°
C. If the temperature exceeds this
range:
The displayed value will be fixed to 50.0
°
C if the range cut is set to “yES”.
The temperature will be displayed within 55.0
°
C when the range cut is set to “no”.
Refer to “ Setting of a hidden area outside of measurement range (display range cut) ” (page 32).
Meas mode
Contact output check setup screen
(display example when r1 and AL are set)
Control valu setup screen
Delay time setup screen
Changes
the value
Determines the value
Cancels the value
Moving average count setup screen
Changes
the value
Determines the value
Cancels the value
Calibration data check screen
Zero calibration
value display
Display for the number of days
elapsed since the latest calibration
Automatic calibration
wait time display
to
Power ON
Displayed for a few seconds
Meas mode