Data
66
J
Cathode electrode
J
Filter housing
J
Flow switch
J
Flow meter
*3: Measurement cannot be performed correctly when the electrical conductivity falls below 8 mS/m (80
μ
S/cm). Be
sure to perform measurement in the above conditions.
*4: If the electrode is exposed to the direct sunlight, the anode and cathode electrode could deteriorate. Install the
instrument where it is not exposed to direct sunlight (indoors).
*5: If a reducing agent or oxidizer exists in a sample, it affects the reading. Do not use a sample containing a reducing
agent such as thiosulfate or sulfite, or oxidizer such as O
3
, hydrogen sulfide or chlorine dioxide, as measurement
cannot be performed correctly.
Product name
RA-30K
Model
RA-30K
Material of electrode
Au
Material of liquid contact section
PPO, EPDM
Operating temperature
0
°
C to 45
°
C
Mass
Approx. 1.8 g
Product name
Filter unit for residual chlorine meter
Model
RA-F
Material of liquid contact section
PVC, EPDM, PP, AS
Operating temperature
0
°
C to 50
°
C (no freezing)
Primary pressure
0 MPa to 0.5 MPa
Piping connection
sample inlet and outlet Union TS16A
Mass
Approx. 3.0 g
Product name
Flow switch for residual chlorine meter
Model
RA-RS30
Setting flow rate
0.3 L/min (ON when the flow rate increases, OFF when the flow rate decreases)
Material of liquid contact section
PVC, EPDM, PP
Operating temperature
0
°
C to 50
°
C (no freezing)
Primary pressure
0 MPa to 0.5 MPa
Installation method
Horizontally (with the switch case located on the upper side)
Piping connection
sample inlet and outlet Union TS16A
Mass
Approx. 0.17 g
Product name
Flow meter for residual chlorine meter
Model
RA-FM30
Setting flow rate
0.1 L/min to 0.5 L/min
Material of liquid contact section
PVC, EPDM, PMMA
Operating temperature
0
°
C to 50
°
C (no freezing)
Primary pressure
0 MPa to 0.5 MPa
Piping connection
sample inlet and outlet
φ
8 mm
×
φ
6 mm tube
Mass
Approx. 0.5 kg