Sensepoint XCD RFD Technical Manual
SPXCDHMRFEN Issue 4_08-2016
74
22
Pentane
3*
23
Propane
5*
24
Propan-2-ol
4*
25
Styrene
2*
26
Tetra hydrafuran
4*
27
Toluene
3*
28
Triethylamine
3*
29
Xylene
2*
Table 14 . Star Rating of Gases
To cross calibrate the Sensepoint XCD flammable gas detector:
(1) Obtain the star rating for both the calibration test gas and the gas to be detected from
Table 14
(2) These values may then be used in Table 15 to obtain the required calibration span
setting when a 50% LEL test gas is applied to the detector.
*Rating of
*Rating of Gas to be Detected
Calibration
Gas
8* 7* 6* 5* 4* 3* 2* 1*
8*
50
62
76
95
- - - -
7*
40
50
61
76
- - - -
6*
33 41 50
62
78 - - -
5*
26 33 40 50 63 79 - -
4*
- 26 32 40 50 63 80 -
3*
- - 26 32 40 50 64 81
2*
- - - 25 31 39 50 64
1*
- - - - 25 31 39 50
Note: These settings must only be used with a calibration gas concentration of 50% LEL.
Table 15. Calibration span setting
(3) If a sensor is to be used to detect a gas other than that for which it was calibrated and
there is no intention to use an equivalent calibration gas to re-calibrate the sensor, then
the required correction factor may be obtained from Table 16.
The reading shown on the gas detector controller or transmitter display should be
multiplied by this number in order to obtain a more accurate gas concentration result.
Sensor calibrated to
Sensor used to detect
detect
8* 7* 6* 5* 4* 3* 2* 1*
8*
1.00 1.24 1.52 1.89 2.37 2.98 3.78 4.83
7*
0.81 1.00 1.23 1.53 1.92 2.40 3.05 3.90
6*
0.66 0.81 1.00 1.24 1.56 1.96 2.49 3.17
5*
0.53 0.66 0.80 1.00 1.25 1.58 2.00 2.55
4*
0.42 0.52 0.54 0.80 1.00 1.26 1.60 2.03
3*
0.34 0.42 0.51 0.64 0.80 1.00 1.27 1.62
2*
0.26 0.33 0.40 0.50 0.63 0.79 1.00 1.28
1*
0.21 0.26 0.32 0.39 0.49 0.62 0.78 1.00
Table 16. Correction factors
Summary of Contents for Sensepoint XCD RFD
Page 1: ...Technical Handbook Sensepoint XCD RFD Remote Flammable Detector ...
Page 65: ...Sensepoint XCD RFD Technical Manual SPXCDHMRFEN Issue 4_08 2016 65 China PA Certification ...
Page 66: ...Sensepoint XCD RFD Technical Manual SPXCDHMRFEN Issue 4_08 2016 66 18 2 Korea KTL ...
Page 68: ...Sensepoint XCD RFD Technical Manual SPXCDHMRFEN Issue 4_08 2016 68 ATEX for Sensor ...
Page 70: ...Sensepoint XCD RFD Technical Manual SPXCDHMRFEN Issue 4_08 2016 70 IEC Ex for Sensor ...
Page 79: ...www norrscope com ...