Sensepoint XCD RFD Technical Manual
SPXCDHMRFEN Issue 4_08-2016
43
bumP
tESt
PEAK READING
DISPLAYED WITH OUTPUT
INHIBITED ‘
’ ICON AND
PEAK STRING FLASHING.
APPLY BUMP TEST
GAS AND CHECK PEAK
READING ON DISPLAY.
CURRENT GAS
CONCENTRATION
DISPLAYED WITH
OUTPUT INHIBITED ‘
’
ICON FLASHING.
UNIT RETURNS TO
MENU MODE
Forc
Curr
‘Forc’ DISPLAYED WITH ‘
’ ICON. DEFAULT FORCE
CURRENT ‘4.00’ FLASHES.
USE
TO CHANGE TO
REQUIRED mA LEVEL.
UNIT TRANSMITS THE
CURRENT
IF YOU WANT TO EXIT
FROM THIS MENU, USE
ABORT FUNCTION
SEt
ALrm
‘AL1’, ICON DISPLAYED
AND CURRENT ALARM
LEVEL 1 CONCENTRATION
FLASHES. USE
TO
CHANGE TO REQUIRED
CONCENTRATION LEVEL.
‘AL1’ DISPLAYED WITH
‘NonE’, ‘rISE’ OR ‘FALL’
FLASHING. USE
TO SELECT REQUIRED
ALARM ACTION.
‘AL2’, ICON
DISPLAYED
AND CURRENT
ALARM LEVEL 2
CONCENTRATION
FLASHES. USE
TO CHANGE
TO REQUIRED
CONCENTRATION
LEVEL.
‘AL2’ DISPLAYED WITH
‘NonE’, ‘rISE’ OR ‘FALL’
FLASHING. USE
TO SELECT REQUIRED
ALARM ACTION AND USE
‘
’ TO RETURN TO MENU
MODE.
Set
rLY
‘rL1’ DISPLAYED AND
‘AL1’, ‘AL2’, ‘Inht’ or ‘FLt’
FLASHES. USE
TO
CHANGE TO REQUIRED
RELAY TARGET.
‘rL1’ DISPLAYED
AND ‘dEEn’ or ‘Enr9’
FLASHES. USE
TO
CHANGE TO REQUIRED
RELAY ACTION.
‘Rl2’ DISPLAYED AND
‘AL1’, ‘AL2’, ‘Inht’ or
‘FLt’ FLASHES. USE
TO CHANGE TO
REQUIRED RELAY
TARGET.
‘rL2’ DISPLAYED AND
‘dEEn’ or ‘Encr9’ FLASHES.
USE
TO CHANGE
TO REQUIRED RELAY
ACTION.
‘rL3’ DISPLAYED AND
‘AL1’, ‘AL2’, ‘Inht’ or ‘FLt’
FLASHES. USE
TO
CHANGE TO REQUIRED
RELAY TARGET.
‘rL3’ DISPLAYED
AND ‘dEEn’ or ‘Enr9’
FLASHES. USE
TO
CHANGE TO REQUIRED
RELAY ACTION.
UNIT RETURNS TO
MENU MODE
rlY
OPr
‘rLY’, ‘ON’ DISPLAYED AND
CURRENT RELAY-ON TIME
FLASHES. USE
TO
CHANGE TO REQUIRED
RELAY-ON DELAY.
‘rLY’, ‘OFF’ DISPLAYED
AND CURRENT RELAY-
OFF TIME FLASHES.
USE
TO CHANGE
TO REQUIRED RELAY-
OFF DELAY.
‘Ltch’ DISPLAYED
AND ‘YES’ or ‘No’
FLASHES. USE
TO CHANGE TO
REQUIRED LATCH
OPTION.
UNIT RETURNS TO MENU
MODE
Set
Loc
‘Loc’ DISPLAYED WITH
FIRST 4 CHARACTERS OF
THE LOCATION STRING.
USE
TO CHANGE THE
1ST CHARACTER OF
CURRENT LOCATION
STRING. USE
TO MOVE
TO NEXT CHARACTER
AND SET REST OF
STRING. MAXIMUM 12
CHARACTERS CAN BE
SET.
‘Loc’ DISPLAYED NEW
LOCATION STRING. THE
STRING MOVES RIGHT-
TO-LEFT TO SHOW
WHOLE CHARACTERS
TWICE. THEN UNIT
AUTOMATICALLY
RETURNS TO MENU
MODE.
tEm
PUnIt
‘tEmP’ DISPLAYED WITH
‘°C’ OR ‘°F’ FLASHING.
USE
TO CHANGE TO
REQUIRED TEMPERATURE
UNIT.
UNIT RETURNS TO
MENU MODE
tES
tALrm
‘Forc’ DISPLAYED and ’AL1’
FLASHING TO SELECT
TEST-REQURED ALARM.
USE ‘
’ TO FORCE
ALARM1.
GAS NAME, CURRENT
GAS CONCENTRATION
DISPLAYED AND ‘
’ ICON AND AL ICON
FLASHE.
UNIT RETURNS TO
MENU MODE
9uIt
QUITS MENU MODE AND
RETURNS TO MONITORING
MODE
OK
OK
OK
OK
1
Refer to section 10.1 for detailed zero and span calibration procedures. Re-calibrate the detector if left un-powered for
periods in excess of 24 hours.
2
The calibration due warning counter is automatically reset after a successful calibration.
Table 12. Configuration Mode Operation
Summary of Contents for Sensepoint XCD RFD
Page 1: ...Technical Handbook Sensepoint XCD RFD Remote Flammable Detector ...
Page 65: ...Sensepoint XCD RFD Technical Manual SPXCDHMRFEN Issue 4_08 2016 65 China PA Certification ...
Page 66: ...Sensepoint XCD RFD Technical Manual SPXCDHMRFEN Issue 4_08 2016 66 18 2 Korea KTL ...
Page 68: ...Sensepoint XCD RFD Technical Manual SPXCDHMRFEN Issue 4_08 2016 68 ATEX for Sensor ...
Page 70: ...Sensepoint XCD RFD Technical Manual SPXCDHMRFEN Issue 4_08 2016 70 IEC Ex for Sensor ...
Page 79: ...www norrscope com ...