Mini-AT User Guide
2-Point Temperature Calibration (Items Required)
- Mini-AT
- A cold temperature source (ice bath)
- A hot temperature source (heated water, not to exceed 170 degrees F.)
- Temperature Reference (calibrated thermometer)
- Computer (IBM Compatible)
- Mini-AT Link (or MasterLink32) Software
- I/O cable, p/n 40-1629
2-Point Temperature Calibration (Procedure)
Refer to figure 21 when temperature calibrating the Mini-AT. Since the two-point calibration is mostly a software
function, references to the appropriate sections in the Mini-AT Link software are included.
1) Prepare an agitated ice bath and a container of heated water.
2) Using serial cable p/n 40-1629, connect from the computer's serial port to the Mini-AT’s RS-232
port.
Fig. 21
2-Point Temperature Calibration
Ice Bath
Heated
Water
Mini-AT
Mini-AT
Link
Calibrated
Thermometer
Calibrated
Thermometer
Temperature
Probe
RS-232 Port
Back View
Pressure Transducer
Case Connector
I/O Cable
p/n 40-1629
Summary of Contents for Mini-AT
Page 2: ...2 www honeywell com ...
Page 138: ...138 www honeywell com Mini AT User Guide Mini AT Door ...
Page 140: ...140 www honeywell com Mini AT User Guide Meter Mount Mini AT ...
Page 142: ...142 www honeywell com Mini AT User Guide Wall Mount Mini AT ...
Page 144: ...144 www honeywell com Page 144 Mini AT User Guide ...
Page 145: ...145 Page 145 Mini AT User Guide ...
Page 146: ...146 www honeywell com Page 146 Mini AT User Guide ...
Page 147: ...147 Page 147 Mini AT User Guide ...
Page 148: ...148 www honeywell com Page 148 Mini AT User Guide ...
Page 149: ...149 Page 149 Mini AT User Guide ...
Page 150: ...150 www honeywell com Page 150 Mini AT User Guide ...