52
Bar Code Calibration Procedure without Remote Display
7. Make sure there is no load on the scale platter. The blue LED will be used to indicate the current step in
the calibration process.
8. Calibrate a zero load. The blue LED will blink once periodically. Wait 8 to 10 seconds for scale stability
then press the beeper volume switch one time. The beeper will beep 1 time indicating that the Cal 1 value
has been stored.
Figure 53.
9. Place a half load on the center of the scale platter. The blue LED will blink twice periodically. Wait for
scale stability, and then press the beeper volume switch once. The beeper will beep 2 times indicating that
the Cal 2 value has been stored.
Figure 54.
10. Add an additional half load to the existing half to simulate a full load (center the entire load). The blue LED
will blink three times periodically. Wait for scale stability, and then press the beeper volume switch once.
The beeper will beep 3 times indicating that the Cal 3 value has been stored.
Figure 55.