ACM 150
50
Exhaust
The ACM 150 monitor exhausts the air it samples through an exhaust pipe located on
the top. Connect this to an exhaust duct with no flow restriction that is compatible
with the types of gases monitored by the ACM 150 monitor. The requirements are:
Flow Rate
12 scfm (340 lpm)
Connection
1-1/2 inch male pipe thread (per
Figure 25
)
Figure 25. Nitrogen, CDA and exhaust connections
1.
Exhaust Pipe
2.
Nitrogen
3.
Air (CDA)
NOTE
Vent the monitor’s exhaust properly in accordance with site or legislated
safety policies governing the gases being monitored, workplace safety, and
environmental requirements