ACM 150
121
Internal gas cell filter
The gas cell filter is part of the mechanical pump assembly,
Figure 12
, located in the
lower pump cabinet. Replace it according to the recommended schedule or more
frequently if required. The ACM 150 monitor should be taken off-line for this service.
If not, expect to generate a Service Status message when temporarily interrupting
the sample flow.
If flow through the filter is over-restricted, the pressure during the sampling cycles
may be inadequate and create a Service Status message. Follow the test procedure,
Cycle valve tests
, and replace the filter if the test indicates a flow restriction.
External in-line filters
These filters are located at or near each sample line inlet. Replace filters according
to the recommended schedule or more frequently if required. You do not have to take
the ACM 150 monitor off-line when replacing filters, because flow is not interrupted.
You can replace all of the filters on the same schedule, or you can replace them
in phases, e.g. 25% every three months. Since sample points are often added in
phases, you might use the same phase interval for filter replacement. It is time
consuming to locate, access and replace the in-line filters.
If flow through one or more filters is over-restricted, the vacuum on those lines
should increase when they are sampled and create a Service Status message. Follow
the test procedure,
Cycle valve tests
, and replace the filter if the test indicates a flow
restriction.
NOTE
Avoid unnecessary downtime. Do not wait until a Service Status message is
generated before replacing filters. Replace filters on an established schedule.
You can modify the schedule based on actual experience at your facility.
Visual inspection
You should visually inspect the gas cell filter and in-line filters during each service.
The filter housing is white and semi-transparent when new. If the housing appears
clean and, if testing indicates there is no flow restriction, you can postpone
replacement until the next service.
If the filter appears dark and or deeply discolored, replace it. In-line filters on dirty
and/or corrosive sample points appear this way much sooner than filters on clean
points. Increase the replacement frequency of filters that foul ahead of schedule.
Light discoloring is normal and not a reason to replace the filter before its scheduled
replacement date.
Servicing the pumps
Self-monitoring and manual testing
The ACM 150 monitor constantly monitors its vacuum pumps. If either venturi pump
fails to provide adequate vacuum, the Backup Exhaust Pump turns on automatically
and a Service Status message notifies the user. A message is also generated if the
Boost Pump fails to provide adequate flow, however, there is no backup for the
Boost Pump.
Summary of Contents for ACM 150
Page 1: ...ACM 150 Air Composition Monitor Guide to Installation Operation and Maintenance ...
Page 2: ......
Page 15: ...ACM 150 15 Figure 1 ACM 150 outside labels ...
Page 16: ...ACM 150 16 Figure 2 ACM 150 inside labels ...
Page 29: ...ACM 150 29 Figure 15 Relay output board ...
Page 59: ...ACM 150 59 Figure 30 Installation diagram ...
Page 152: ...ACM 150 152 ...
Page 153: ......
Page 154: ......
Page 155: ......
Page 156: ...ACM 150 Air Composition Monitor Revision 2 ECO HAA150063 December 2015 Honeywell Analytics ...