The second pump is the Gas Cell Venturi Pump. It draws spent sample out of the gas
cell during the Evacuation mode.
The advantages of the air-driven venturi pump are:
•
no scheduled PM service because parts do not wear out
•
generates virtually no noise or heat
Should the supply of air be interrupted, a mechanical pump turns on and takes over
automatically.
Figure 11. Venturi pump assembly
The venturi pumps are installed on a mounting plate. The check valves allow the
venturi pumps and the mechanical Backup Exhaust Pump to be pneumatically
coupled to common bypass and gas cell vacuum lines. This allows either one
to provide the suction to draw the flow. The mechanical pump switches on
automatically when the venturi pumps fail to provide adequate vacuum. The check
valves permit flow in one direction only. This prevents the pump in use from drawing
on the idle pump.
Mechanical pumps
The pumps are isolated in the bottom section of the ACM 150 monitor to muffle the
noise and to isolate vibration and heat. Cooling fans in the bottom section of the
cabinet direct air flow across the pumps and their motors. The pumps are accessed
by removing the front panel. They are mounted with vibration eliminators on a
common plate that slides out for service.
Summary of Contents for ACM 150
Page 1: ...ACM 150 Air Composition Monitor Guide to Installation Operation and Maintenance ...
Page 2: ......
Page 15: ...ACM 150 15 Figure 1 ACM 150 outside labels ...
Page 16: ...ACM 150 16 Figure 2 ACM 150 inside labels ...
Page 29: ...ACM 150 29 Figure 15 Relay output board ...
Page 59: ...ACM 150 59 Figure 30 Installation diagram ...
Page 152: ...ACM 150 152 ...
Page 153: ......
Page 154: ......
Page 155: ......
Page 156: ...ACM 150 Air Composition Monitor Revision 2 ECO HAA150063 December 2015 Honeywell Analytics ...