Vacuum sampling system
The ACM 150 monitor samples up to 40 areas, which are scanned and analyzed in
sequence, as described in Air sampling system. The main assemblies are:
Bypass Venturi Pump
Draws flow through sample tubes and bypasses the
flow to exhaust to provide a constant flow of fresh
sample to the monitor
Gas Cell Venturi Pump
Evacuates the sampled air from the gas cell after it is
analyzed
Boost Pump (mechanical)
Fills the evacuated gas cell with fresh sampled air for
each analysis cycle
Sample Valve Assembly
Valves installed in the 10-port sample manifolds
(up to 40) that switch in sequence as the ACM 150
monitor scans the sample areas
Cycle Valve Assembly
Group of 6 valves that control the flow through the
gas cell
Backup Exhaust Pump
Normally idle, this mechanical pump backs up the
venturi pumps to provide operation in the absence of
CDA
Available outputs
The following output relays are dry contact, double-pole/double-throw (DPDT), rated
for 3 amps @ 30 VDC
Programmable Output
Relays (80)
Default programmable relay settings
•
Relays 1-40 activate for any gas at the Alarm 1
level set point for areas 1-40
•
Relays 41-80 activate for any gas at the Alarm 2
level set point for areas 41-80
The default settings for each relay may be changed by
the user to activate
•
At the Alarm 1 or Alarm 2 level
•
For any gas, for all gases or any selected group of
gases
•
At any area, all areas or any selected group of
areas
Additional Output Relays (6)
•
Power fail
•
Watchdog timer
•
Manual Mode
•
General malfunction
•
General warning
•
General alarm
Data interface and protocol
Network
Ethernet
Protocols
TCP/IP webserver
Optional Driver
OPC (OLE for Process Control)
Other communication networks are available. Contact Honeywell Analytics for more
information.
Summary of Contents for ACM 150
Page 1: ...ACM 150 Air Composition Monitor Guide to Installation Operation and Maintenance ...
Page 2: ......
Page 15: ...ACM 150 15 Figure 1 ACM 150 outside labels ...
Page 16: ...ACM 150 16 Figure 2 ACM 150 inside labels ...
Page 29: ...ACM 150 29 Figure 15 Relay output board ...
Page 59: ...ACM 150 59 Figure 30 Installation diagram ...
Page 152: ...ACM 150 152 ...
Page 153: ......
Page 154: ......
Page 155: ......
Page 156: ...ACM 150 Air Composition Monitor Revision 2 ECO HAA150063 December 2015 Honeywell Analytics ...