After the polynomial has been calculated it is drawn on the screen and determines the
different absorption
measurement results
. The results shown in blue in the example above
are:
Maximum torque
The absolute maximum torque of the measurement. The
result was 6418 mNm.
Proc.A (TLS)
The intersection between the TLS value (5000 mNm) and the
polynomial. The result was 65.3 ml/100g here.
Proc.B (70 %)
The intersection between 70 % of the maximum torque and
the polynomial. The result was 64.7 ml/100g.
The test conditions in our example were:
Measurement Threshold
1 000 (mNm)
Measurement End:
20 (s)
Measurement Torque limit:
10 000 (mNm)
Polynomial Start:
20 (%)
Polynomial End:
10 (s)
Refer to
to get more information about the different standards used.