C-6
Moving and Installing the System
7. Place tape to secure items inside the Sera Handlng Area of the AP3600. Install Fixing Plates. See Figure C-7.
FIGURE C-7. Place Tape to Secure Items
8. Remove the reagent bottles from the right side of the instrument.
9. Remove the Wash Buffer, DI, and waste bottles under the instrument.
10. Disconnect any USB devices or network connections to the AP3600 instrument at the Ethernet port or USB connections on the front
of the instrument and under the monitor.
11. Disconnect the power cable from the back of the AP3600 instrument and disconnect the cable from the Bottle Unit.
12. Secure the outside of the AP3600 with tape and bubble wrap. See Figure C-8.
FIGURE C-8. Place Tape and Bubble Wrap to Secure Items
13. Lift the AP3600 Instrument from the bottom of the unit and place it on a clean, flat work table (See Figure C-1) or appropriate sized
cart for moving.
Z Axis of Distribution Nozzle and
X and Y Axis Fixing Plates
Sample Cup Carousel
Left Door and Top Door
Rear Right Carousel