OPTIGEN® AP3600™ AUTOMATED INSTRUMENT USER MANUAL
Specifications and Ordering Information
Physical Specifications
• Dimensions: W x H x D = 1140 x 630 x 760 mm
• Weight: 170 kg (375 lbs.)
Environmental Requirements
• Operating Temperature: 15 – 33 °C
• Operating Humidity: 10 – 85%, non-condensing
• Altitude: Less than 2000 m
• Over-voltage category II
Power Requirements
• Input Voltage: 100 VAC to 240 VAC ± 10%
• Input Frequency: 50 /60 Hz ± 4%
• Power Consumption: 500 VA
• AP3600 Instrument Power Cord Voltage/Current:
• USA and Canada (110-120 VAC): 125V 10A power cord. Manufacturer: KAWASAKI ELECTRIC WIRE CO.,LTD. (KDK);
Model: KP-30 SVT3/18AWG.
• Countries using 200-240 VAC: 250V 7A cord. Acquire power cord locally.
WARNING: Do not use an inadequately rated power cord with the AP3600 Instrument. This may damage the instrument.
Processing Specifications
• Number of Pettes Processed: Maximum of 50 Pettes per run
• Processing Time: One run completes in 5.5 hours (excludes automatic waste solution operation)
Ordering Information
To order AP3600 accessories or OPTIGEN
®
Allergy Testing Kits, contact Hitachi Chemical Diagnostics or an authorized representative.
Hitachi Chemical Diagnostics, Inc.
Tel: +1 650 961 5501
Fax: + 1 650 969 2745
E-mail: [email protected]
B-1
Specifications and Ordering Information
Summary of Contents for OPTIGEN AP3600
Page 1: ...OPTIGEN AP3600 AUTOMATED INSTRUMENT USER MANUAL ...
Page 2: ...OPTIGEN AP3600 Automated Instrument UserManual Rx Only ...
Page 6: ...OPTIGEN AP3600 AUTOMATED INSTRUMENT USER MANUAL Glossary E 1 v Table of Contents ...
Page 14: ...1 6 Safety This page intentionally left blank ...
Page 54: ...8 8 Utilities Screen This page intentionally left blank ...
Page 61: ...9 7 System Settings Screen This page intentionally left blank ...
Page 62: ......
Page 66: ...10 4 End of Run This page intentionally left blank ...
Page 82: ...D 4 Maintenance This page intentionally left blank ...