6.2 Basic Specifications
50
6.2 Basic Specifications
Number of channels
8 (parallel and simultaneous measurement)
Measurement Method
Applies voltage to the measurement target and measures current.
Applied voltage
Supplied from an external power supply (Input to the voltage input connector on the
rear panel.)
Input resistance of each
current meter
1 k
Ω
±10%
Input/Output terminals
Current input terminals (Front panel) :HIOKI proprietary input connector for IR meters
Voltage output terminals (Front panel):Safety terminal (Red-colored)
Voltage input terminals (Rear panel) :Special round connector
Maximum input voltage:1000VDC (between each terminal)
1000VDC (between the ground and terminals)
Setup method and
operation
GP-IB Interface
RS-232C Interface
EXT I/O
(No setup available on the instrument front panel except settings for GP-IB address)
LED indicator
For “Power indication” and “Trigger indication”
LCD screen
3-page layout
P1: For “measurement values” and “current measurement ranges”
P2: For “Contact-check measurement values”, “Results of contact-check” and “Mea-
surement voltage values”
P3: For “Measurement speed”, “Trigger Delay”, “Averaging settings” and “GP-IB ad-
dress”
Recommended
power supply
HIOKI Model SM7860 series Power Source Unit