5.2 Timing Chart
44
Each signal level indicates a corresponding voltage level.
5.2 Timing Chart
TRIG
INDEX
EOM
NO CONTACT
1 to 8
HI 1 to 8
IN 1 to 8
LO1 to 8
t1
t2
t3
t4
t5
t6
t7
Contact-
check
T delay
Measur
ement
Previous measurement results
New measurement results
Normal measurement
Timing Chart Interval Descriptions
Interval
Description
Duration
t1
Trigger pulse width (Low time)
100
μ
s or more
t2
Trigger OFF (Hi time)
100
μ
s or more
t3
INDEX, EOM delay time
200
μ
s or less
t4
T index (Measurement time)
Within (set measurement time + T delay)
t5
T eom
Within (T index + 500
μ
s)
t6
Trigger setup time
Display ON : 30 ms or more
Display OFF: 1 ms or more
t7
NO CONTACT delay time
3 ms or less