7.2 Replacing the Power Fuse
60
7.2 Replacing the Power Fuse
•
To avoid electric shock, turn off the power switch and disconnect the
connection cables before replacing the fuse.
•
Replace the fuse only with one of the specified characteristics and volt-
age and current ratings. Never use unspecified fuses and never use the
instrument after the fuse holder has shorted. This will damage the instru-
ment and cause injury.
Fuse type: 250 V T0.315AL
φ
5 mm × 20 mm, Slo-Blo type
Power inlet
Screwdriver
Fuse holder
Fuse:
φ
5 mm × 20 mm
Replacing the Power Fuse
Rear panel of the instrument
Tools to Prepare: Flat blade screwdriver
1
Turn off the power switch and
disconnect the power cord.
2
Align the flat blade screwdriver
with the fuse holder securing
part of the power inlet and then
remove the fuse holder by
pushing the handle of the
screwdriver toward the oppo-
site side of the unit.
3
Replace the power fuse with a
fuse of the designated rating.
4
Reinsert the fuse holder in the
power inlet.