5.2 Timing Chart
46
Contact-check
C.CHECK
t1
t2
t3
t4
t6
t5
Previous measurement results
New measurement results
INDEX
EOM
NO CONTACT
1 to 8
Timing Chart Interval Descriptions
Interval
Description
Duration
t1
Pulse width (Low time)
100
μ
s or more
t2
INDEX, EOM delay time
400
μ
s or less
t3
T index (Measurement time)
4 ms or less
t4
T eom
4 ms or less
t5
Judgment setup time
100
μ
s or more
t6
Input setup time
2 s or more