4.6 Message List
38
LCD display
LCD
LCD display mode setting
d1 (Display mode: 0 to 1)
0: OFF Display OFF
1: ON Display ON
[Format]
LCD d1
d1: NR1 format
LCD?
LCD display mode query
The contents of responses are the same as the settings.
[Format]
LCD?
[Response] d1
PAG
LCD display page specification
d1 (Page number: 0 to 2)
0: Displays measured value.
1: Displays contact check results.
2: Displays operation conditions.
[Format]
PAG d1
d1: NR1 format
Measurement data
RDT?
Measurement data query
d1 (Format specification: 0 to 2)
0: Fundamental waveform
1: Measured value only
2: Comparison results only
Note: When the comparative measurement function is OFF, no
query is returned even if the
RTD? 2
command is executed.
For more information about the response, see "Measure-
[Format]
RDT? d1
[Response] d1: NR1 format
MTG
Manual trigger
d1 (Format specification: 0 to 2 [may be omitted])
If omitted: No automatic data return
0: Fundamental waveform
1: Measured value only
2: Comparison results only
For more information about the response, see
[Format]
MTG d1
d1: NR1 format
Command
Description
Formats
Summary of Contents for SM7810
Page 19: ...2 5 Turning the Power On and Off 16 ...
Page 51: ...5 3 Internal Circuitry 48 ...
Page 65: ...7 4 Cleaning 62 ...