36
Affixing the Voltage Sensors
3.3
Affixing the Voltage Sensors
Affix the voltage sensors R (1), S (2), and T (3) to the cables respectively of
phase R (1), S (2), and T (3).
Measurement with
the instrument in your
hand
Measurement using the Model Z5020
Magnetic Strap
R S T
R
R
T
T
S
S
(1)(2)(3)
(1)(2)(3)
(1)(2)(3)
R S T
(1)(2)(3)
R
R
T
T
S
S
(1)(2)(3)
(1)(2)(3)
Phase labeling systems
Many labeling systems exist for three-phase systems.
<Examples>
First phase
Second phase
Third phase
R
S
T
L1
L2
L3
A
B
C
U
V
W
Affix voltage sensors R (1), S (2), and T (3) respectively to the first, second,
and third phases of the three-phase circuit.