Troubleshooting
Errors and status codes
The instrument will show error codes and other information,
according to the error and operation status, on the display.
If the instrument is in need of repair, contact your authorized Hioki
distributor or reseller.
Code
Description
Remedy
Err 001
ROM error program
Replace the batteries with fresh
If this fails to resolve the issue, the
instrument is in need of repair.
Err 002
ROM error adjustment
data
Err 005
ADC error
Hardware failure
Err 008
Z3210 communications
error
Poor connection, failure
in Z3210 or hardware
Follow the instruction below.
Connect another Z3210.
If the instrument still shows the
error code, contact your authorized
Hioki distributor or reseller.
APS
→
P.oFF
Shutdown by auto-power
save function
Turn on the instrument again.
bAtt
→
P.oFF
Shutdown due to battery
exhaustion
Replace the batteries with fresh