![Hioki PD3259-50 Instruction Manual Download Page 76](http://html1.mh-extra.com/html/hioki/pd3259-50/pd3259-50_instruction-manual_2137432076.webp)
Voltage Sensors of the PD3259-50
70
6.2 Voltage Sensors of the PD3259-50
Each voltage sensor of the instrument incorporates an internal
electrode (metal plate). When the voltage sensor is affixed to the
measurement target (cable), the minuscule current
I
flows due
to capacitive coupling between the measurement target and the
voltage sensor’s internal electrode.
I
fCV
=
2
π
(1)
I
fCV
=
2
π
: Frequency of measurement target (Hz)
I
fCV
=
2
π
: Capacitance between the measurement target and the voltage
sensor’s internal electrode (F)
I
fCV
=
2
π
: Voltage (AC) between the measurement target and the voltage
sensor’s internal electrode (V)
V
1
V
2
I
fCV
=
2
π
I
fCV
=
2
π
Voltage sensor’s
internal electrode
Voltage
generation
circuit
Measurement
target (cable)
HIOKI PD3259C961-00
Summary of Contents for PD3259-50
Page 2: ...HIOKI PD3259C961 00...
Page 6: ...iv Contents HIOKI PD3259C961 00...
Page 26: ...20 Part Names and Functions HIOKI PD3259C961 00...
Page 38: ...32 Inspecting the Instrument HIOKI PD3259C961 00...
Page 64: ...58 General Specifications Accessories See p 2 Options See p 3 HIOKI PD3259C961 00...
Page 74: ...68 Troubleshooting HIOKI PD3259C961 00...
Page 78: ...Voltage Sensors of the PD3259 50 72 HIOKI PD3259C961 00...
Page 82: ...Index 76 W Wireless Adapter 3 28 52 Wireless communications 28 52 HIOKI PD3259C961 00...
Page 83: ...HIOKI PD3259C961 00...
Page 84: ...HIOKI PD3259C961 00...
Page 85: ...HIOKI PD3259C961 00...