4.7 Start and Stop Operation (MEM)
______________________________________________________________
Trigger
conditions are
met
Measurement
start
Recording data
Measurement
end
Trigger
mode
REPEAT
AUTO
STOP
SINGLE
NOTE
If measurement is aborted after a trigger event occurs, waveform
data up to the abort are retained. However, no data is retained if
measurement is aborted before a trigger event. If the trigger
mode is Continuous or Auto, and waveforms remain from a
previous trigger event, the waveforms are displayed (excluding
maximum recording length).
4.7 Start and Stop Operation (MEM)
Press the
START
button to start measurement, and
LED lights in green. The
START
button activation
condition can be selected. Refer to Section 10.1.1.
When pre-trigger is activated, the trigger is not
enabled until the specified pre-trigger data has been
read.("Pre-trig wait" is displayed.)
"Wait for trig" is displayed until the trigger condition
occurs.
Data recording starts when the trigger conditions are
met. When trigger mode is AUTO, the product waits
for about 1 second for trigger conditions to be met.
After this interval, data recording starts, regardless of
trigger state.
During recording, "Storing" is displayed.
When the Roll mode is enabled, measurement data is
displayed (scrolling) as it is recorded. When automatic
printing is also enabled, data prints at the same time it
is displayed.
Measurement can be aborted by pressing the
STOP
button twice during measurement.
Then the waveform data until the
STOP
button is
pressed is displayed and saved. When the auto saving
is enabled, the data until a forced termination is
automatically saved. (automatic printing will not
execute)
Trigger mode: REP/AUTO
Activates the trigger waiting condition. Input resumes
when the trigger conditions are met.
Trigger mode: SING
Measurement can be aborted by pressing the
STOP
button twice (automatic printing will not execute).
Trigger mode: REP/AUTO
During measurement, pressing the
STOP
button once
records up to the set recording length (waveform
display, printing and auto-save are executed).
Measurement can be aborted by pressing the
STOP
button twice (automatic printing will not execute).