APPENDIX 2
______________________________________________________________________
Appendix 1 Messages
________________________________________________________________
93 Disk full.
Little capacity is available.
Delete files or use new media.
94 Path name error.
Up to 127 characters can be used for a path name.
95 Empty directory name.
Name the directory.
96 Directory already exists.
Another directory of the same name exists.
Saving cannot be performed.
99 Conditions for OVERWRITE
are not satisfied.
Match the product condition and file data
condition (function and time-axis).
201 Set printer paper.
Printer paper has run out. Reload.
202 Set printer lever
The head up/down lever has been left in the up
position.
203 Cannot start.
Start measurement after printing is completed.
205 Invalid (START)
The button pressed is not valid, because
measurement operation is in progress.
207 AUTO RANGE failure.
The auto ranging function has failed. Check the
input signal.
300 Cannot START.
(SYSTEM)
Cannot start measurement from screen displayed
by pressing the SYSTEM button.
345 Cannot set. (AND logic
trigger)
The trigger source "AND" cannot be set between
the logic trigger and analog trigger of rms
function.
346 Cannot set.
(AND analog trigger)
The trigger source "AND" cannot be set between
the logic trigger and analog trigger of rms
function.
347 Invalid. (Pre Trigger)
If the trigger is not set, the pre-trigger is disabled
(0%).
348 Invalid. (V-drop Trigger)
When the time axis range is other than 200
μ
s to
5 ms/DIV, the voltage drop trigger is disabled.
349 Invalid.(Real Trigger)
When the time axis range is other than 10
ms/DIV, the waveform evaluation trigger is
disabled.
382 No waveform data.
There is no waveform data to display on the
Waveform Screen. Data must be input before
processing can start.
397 Out of range. (scaling)
Input the appropriate value.
398 A/B cursor positions
invalid.
Move the cursors to the appropriate position.
503 Communication settings
is incorrect.
Make the communication settings correctly.
504 Communication error
The communications line was disconnected
repeatedly. Turn off the power and turn it on
again.