274
_____________________________________________________________________
14.2 Trigger
______________________________________________________________
Harmonic wave
trigger mode
Single, repeat
Harmonic wave
trigger source
Up to four different harmonic wave triggers can be
selected; trigger conditions can be set separately for
each trigger type. Free run operation if all triggers
are off.
Trigger conditions
Harmonic wave triggers can be ORed.
Harmonic wave trigger sources and external or timer
sources can be ANDed.
Harmonic wave
triggers
Effective value trigger for any harmonic wave
Component ratio trigger for any harmonic wave
Active power value trigger for any harmonic wave
Power component ratio trigger for any harmonic
wave
Power phase angle trigger for any harmonic wave
Total effective value trigger
Total distortion trigger*
Trigger level
resolution
One digit of the analysis display value
One degree (phase angle trigger)
Pre-trigger
0, 5, 10 DIV (time series analysis function)
Trigger timing
Start only
14.2.1 Harmonic Wave Trigger
*: Total distortion-F, total distortion-R